Refurbished Tencor Surfscan Comparison: 6100, 6220 and 6420

Five refurbished Surfscan wafer inspection systems compared side by side: particle sensitivity, laser, illumination angle, haze, SECS/GEM data output, calibration wafers and warranty.

The five refurbished Tencor Surfscan systems we offer (often listed as KT Surfscan or SFS) are unpatterned wafer inspection systems. Each one detects, counts and sizes particles and measures haze on bare silicon wafers up to 200 mm (8 inch). They differ in three main attributes: particle sensitivity (≈ 204 nm down to ≈ 100 nm at a 95% capture rate), laser (red 632 nm HeNe or blue 488 nm solid-state) and factory data output (SECS/GEM optional or installed).

Quick Answer: Which Surfscan Fits Which Need

  • Smallest particles: Surfscan 6420 Ultimate, ≈ 0.10 µm (100 nm), with low-angle (oblique) illumination.
  • SECS/GEM installed with normal-incidence optics: Surfscan 6220 Ultimate, ≈ 0.12 µm (120 nm).
  • Haze measurement to 0.02 ppm and calibration wafers included, with a HeNe laser that matches Surfscan 4500 and 5500 particle data: Surfscan 6100 Ultimate, ≈ 0.16 µm (160 nm).
  • New HeNe laser and brand-new main PC boards, with options added as needed: Surfscan 6100 Enhanced, ≈ 0.16 µm (160 nm).
  • Replacing an obsolete Surfscan 4000, 4500, 5000 or 5500: Surfscan 6100 Upgraded, ≈ 0.204 µm (204 nm).

Surfscan Comparison Table: All Five Refurbished Models

Every row below matches the specification on each product page. Scroll sideways on a phone to see all five models.

FeatureSurfscan 6100 UpgradedSurfscan 6100 EnhancedSurfscan 6100 UltimateSurfscan 6220 UltimateSurfscan 6420 Ultimate
Particle detection on bare silicon≈ 0.204 µm (204 nm) at 95% capture rate≈ 0.16 µm (160 nm) at 95% capture rate≈ 0.16 µm (160 nm) at 95% capture rate≈ 0.12 µm (120 nm) at 95% capture rate≈ 0.100 µm (100 nm) at 95% capture rate
Haze (surface roughness) measurements / resolutionYes, customer calibratesYes, customer calibratesHaze down to 0.02 ppm, resolution @ 0.002 ppmHaze down to 0.02 ppm, resolution @ 0.002 ppmHaze down to 0.02 ppm, resolution @ 0.002 ppm
Laser and scanRed 632 nm HeNe, gas tube laserNew, red 632 nm HeNe, gas tube laserNew, red 632 nm HeNe, gas tube laserNew, blue, solid state 488 nm laserNew, blue, solid state 488 nm laser
Optical detectorPhoto-multiplier tube (PMT)Photo-multiplier tube (PMT)Photo-multiplier tube (PMT)Photo-multiplier tube (PMT)Photo-multiplier tube (PMT)
Incidence angleNormal-incidence illumination for unpatterned silicon wafer defect detection down to 204 nm & haze (surface roughness) measurementsNormal-incidence illumination for unpatterned silicon wafer defect detection down to 160 nm & haze (surface roughness) measurementsNormal-incidence illumination for unpatterned silicon wafer defect detection down to 160 nm & haze (surface roughness) measurementsNormal-incidence illumination for unpatterned silicon wafer defect detection down to 120 nm & haze (surface roughness) measurementsLow-angle (oblique) incidence illumination for unpatterned silicon wafer defect detection down to 100 nm & haze (surface roughness) measurements
Main PC boardsChecked for system operation, engineering updatesBrand-new main PC boards supporting original designBrand-new, improved main PC boards, optional SECS/GEM data outputBrand-new, improved main PC boards, including SECS/GEM data outputBrand-new, improved main PC boards, including SECS/GEM data output
SECS/GEM data outputOptionalOptionalOptionalYes, SECS/GEM data protocol installedYes, SECS/GEM data protocol installed
Hard drivesSolid stateSolid stateSolid stateSolid stateSolid state
Optical mirrors cleaned and adjustedYes, optical system cleaned and adjusted for red 632.8 nm laserYes, optical system cleaned and adjusted for red 632.8 nm laserYes, optical system cleaned and adjusted for red 632.8 nm laserYes, optical system cleaned and adjusted for blue 488 nm laserYes, optical system cleaned and adjusted for blue 488 nm laser
Calibration wafer standardsOptional, 3 full-deposition calibration wafer standards, 220 nm, 500 nm, 1019 nmOptional, 3 full-deposition calibration wafer standards, 204 nm, 500 nm, 1019 nmIncluded, 3 full-deposition calibration wafer standards, 204 nm, 500 nm, 1019 nmIncluded, 4 full-deposition calibration wafer standards, 126 nm, 304 nm, 700 nm, 1019 nmIncluded, 4 full-deposition calibration wafer standards, 100 nm, 304 nm, 700 nm, 1019 nm
Installation and calibrationOptional onsite install, system check and size calibrationOptional onsite install, system check and size calibrationOptional onsite install, system check and size calibrationOptional onsite install, system check and size calibrationOptional onsite install, system check and size calibration
Windows print serverOptionalYesYesYesYes
Warranty6 months optional, travel not included6 months optional, travel not included6 months included; extended optional, travel not included1 year included; extended optional, travel not included1 year included; extended optional, travel not included
AvailabilityJanuary 2027January 20271st Quarter 20271st Quarter 20272nd Quarter 2027

Key Differences Between the Surfscan 6100, 6220 and 6420

Particle sensitivity

Detection at a 95% capture rate on bare silicon is ≈ 204 nm on the 6100 Upgraded, ≈ 160 nm on the 6100 Enhanced and 6100 Ultimate, ≈ 120 nm on the 6220 Ultimate and ≈ 100 nm on the 6420 Ultimate.

Laser: red 632 nm HeNe vs blue 488 nm solid-state

The Surfscan 6000 series was originally built with a 30 mW argon gas laser. The refurbished 6100 models use a red 632 nm HeNe gas tube laser (new on the Enhanced and Ultimate), which keeps particle data in line with Surfscan 4500 and 5500 tools. The 6220 and 6420 Ultimate use a new blue 488 nm solid-state laser. For an existing tool that still runs an argon laser, see the Surfscan Solid State Laser Upgrade.

Illumination angle: normal vs low-angle (oblique)

The three 6100 models and the 6220 Ultimate use normal-incidence illumination. The 6420 Ultimate uses low-angle (oblique) incidence. Its manufacturer has described the Surfscan 6420 as an industry standard for particle sensitivity on rough films, particularly metal layers.

Haze (surface roughness) measurement

The three Ultimate models measure haze down to 0.02 ppm with 0.002 ppm resolution. The 6100 Upgraded and Enhanced also measure haze, but the customer calibrates it.

SECS/GEM data output

SECS/GEM is installed on the 6220 and 6420 Ultimate for direct reporting to your factory HOST computer. It is optional on the 6100 Upgraded, Enhanced and Ultimate.

Calibration wafer standards

The 6100 Ultimate includes three full-deposition calibration wafer standards (204, 500 and 1019 nm). The 6220 Ultimate includes four (126, 304, 700 and 1019 nm) and the 6420 Ultimate four (100, 304, 700 and 1019 nm). They are optional on the 6100 Upgraded (220, 500 and 1019 nm) and the 6100 Enhanced (204, 500 and 1019 nm). Additional sizes are available as PSL calibration wafer standards.

Warranty and availability

The 6220 and 6420 Ultimate include a 1-year warranty and the 6100 Ultimate a 6-month warranty, with extended warranty optional. On the 6100 Upgraded and Enhanced, a 6-month warranty is optional. Travel is not included. The 6100 Upgraded and Enhanced are available in January 2027, the 6100 Ultimate and 6220 Ultimate in the first quarter of 2027, and the 6420 Ultimate in the second quarter of 2027.

What all five have in common

Every model uses a photomultiplier tube (PMT) detector, solid-state drives and an optical system cleaned and adjusted for its laser. Onsite installation, system check and size calibration are optional on all five.

Surfscan 6200 Series vs 6400 Series: 6220 Ultimate or 6420 Ultimate?

The 6220 belongs to the Surfscan 6200 series and the 6420 to the 6400 series. Both refurbished Ultimate systems share a new blue 488 nm solid-state laser, brand-new, improved main PC boards with SECS/GEM installed, haze measurement to 0.02 ppm, four full-deposition calibration wafers and a 1-year warranty. They differ in particle sensitivity, illumination angle, calibration wafer sizes and availability:

FeatureSurfscan 6220 UltimateSurfscan 6420 Ultimate
Particle detection on bare silicon≈ 0.12 µm (120 nm) at 95% capture rate≈ 0.100 µm (100 nm) at 95% capture rate
Incidence angleNormal-incidence illumination for unpatterned silicon wafer defect detection down to 120 nm & haze (surface roughness) measurementsLow-angle (oblique) incidence illumination for unpatterned silicon wafer defect detection down to 100 nm & haze (surface roughness) measurements
Calibration wafer standardsIncluded, 4 full-deposition calibration wafer standards, 126 nm, 304 nm, 700 nm, 1019 nmIncluded, 4 full-deposition calibration wafer standards, 100 nm, 304 nm, 700 nm, 1019 nm
Availability1st Quarter 20272nd Quarter 2027

Surfscan 6100 Upgraded vs Enhanced vs Ultimate

All three are refurbished Surfscan 6100 systems with a red 632 nm HeNe laser. What changes is the level of refurbishment: whether the laser and main PC boards are new, the haze specification and what is included.

Feature6100 Upgraded6100 Enhanced6100 Ultimate
Particle detection on bare silicon≈ 0.204 µm (204 nm) at 95% capture rate≈ 0.16 µm (160 nm) at 95% capture rate≈ 0.16 µm (160 nm) at 95% capture rate
Laser and scanRed 632 nm HeNe, gas tube laserNew, red 632 nm HeNe, gas tube laserNew, red 632 nm HeNe, gas tube laser
Main PC boardsChecked for system operation, engineering updatesBrand-new main PC boards supporting original designBrand-new, improved main PC boards, optional SECS/GEM data output
Haze (surface roughness) measurements / resolutionYes, customer calibratesYes, customer calibratesHaze down to 0.02 ppm, resolution @ 0.002 ppm
Calibration wafer standardsOptional, 3 full-deposition calibration wafer standards, 220 nm, 500 nm, 1019 nmOptional, 3 full-deposition calibration wafer standards, 204 nm, 500 nm, 1019 nmIncluded, 3 full-deposition calibration wafer standards, 204 nm, 500 nm, 1019 nm
Windows print serverOptionalYesYes
Warranty6 months optional, travel not included6 months optional, travel not included6 months included; extended optional, travel not included

Which Refurbished Surfscan Should You Choose?

You need the smallest particle sensitivity

Choose the Surfscan 6420 Ultimate: ≈ 0.10 µm (100 nm) at a 95% capture rate with low-angle (oblique) illumination.

Your fab needs SECS/GEM host reporting

The 6220 Ultimate and 6420 Ultimate ship with SECS/GEM installed. On the 6100 models it is an option.

You are replacing a Surfscan 4000, 4500, 5000 or 5500

Choose a Surfscan 6100. Its red 632 nm HeNe laser keeps particle data in line with Surfscan 4500 and 5500 tools. The 6100 Upgraded is specified as a replacement for these tools.

You want haze measurement and calibration wafers included

Choose an Ultimate model: 6100, 6220 or 6420. All three measure haze to 0.02 ppm and include full-deposition calibration wafer standards.

You want a new laser and new main boards on a 6100

Choose the 6100 Enhanced, with a new HeNe laser and brand-new main PC boards. Calibration wafers, warranty and installation can be added.

You already own a Surfscan 6100, 6200 or 6400 series tool

Keep it running with a solid-state laser upgrade, a CPU upgrade kit, a full refurbishment service or an annual parts contract.

Surfscan Comparison FAQ

What is the difference between the Surfscan 6220 and 6420?

The refurbished 6420 Ultimate detects particles down to ≈ 0.10 µm (100 nm) with low-angle (oblique) illumination. The 6220 Ultimate reaches ≈ 0.12 µm (120 nm) with normal-incidence illumination. Both have a new 488 nm solid-state laser, SECS/GEM installed, haze measurement to 0.02 ppm and a 1-year warranty.

Which refurbished Surfscan has the best particle sensitivity?

The Surfscan 6420 Ultimate, which detects particles down to ≈ 0.10 µm (100 nm) at a 95% capture rate on bare silicon.

Which Surfscan models include SECS/GEM data output?

The Surfscan 6220 Ultimate and 6420 Ultimate have the SECS/GEM data protocol installed. It is optional on the Surfscan 6100 Upgraded, Enhanced and Ultimate.

Which Surfscan uses a HeNe laser and which uses a 488 nm laser?

The refurbished Surfscan 6100 models use a red 632 nm HeNe gas tube laser. The 6220 and 6420 Ultimate use a new blue 488 nm solid-state laser.

Which Surfscan replaces a Surfscan 4500 or 5500?

The Surfscan 6100 replaces obsolete Tencor Surfscan 4000, 4500, 5000 and 5500 tools, and its red 632 nm HeNe laser keeps particle data in line with the Surfscan 4500 and 5500.

Is a used Surfscan the same as a refurbished Surfscan?

Every system we offer is a used Tencor Surfscan that is refurbished before sale. What is replaced depends on the model: the 6100 Upgraded keeps its checked main PC boards, while the Enhanced and Ultimate models get brand-new main PC boards and a new laser.

Browse Refurbished Surfscan Systems

See all models, upgrades and service plans in refurbished Tencor Surfscan wafer inspection systems, or calibrate any Surfscan with NIST-traceable PSL wafer standards.